Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 20 de 26
Filtrar
1.
World J Urol ; 42(1): 298, 2024 May 06.
Artigo em Inglês | MEDLINE | ID: mdl-38709327

RESUMO

PURPOSE: The aim of the study was to evaluate illumination properties in an in-vitro kidney calyx model in saline. DESIGN AND METHODS: We evaluated a series of contemporary flexible ureteroscopes including the Storz Flex-Xc and Flex-X2s, Olympus V3 and P7, Pusen 7.5F and 9.2F, as well as OTU WiScope using a 3D-printed closed pink kidney calyx model, submerged in saline. A spectrometer was used for illuminance and color temperature measurements at different openings located at center (direct light), 45° (direct and indirect light) and 90°(indirect light) to the axis of the scope. RESULTS: Maximum illuminance was at the center opening for all scopes (range: 284 to 12,058 lx at 50% brightness and 454 to 11,871 lx at 100% brightness settings). The scope with the highest center illuminance (Flex-Xc) was 26 times superior to the scope with the lowest illuminance (Pusen 7.5Fr) at 100% brightness setting. For each scope, there was a peripheral illuminance drop ranging from - 43 to - 92% at 50% brightness and - 43% to - 88% at 100% brightness settings, respectively (all p < 0.01). Highest drop was for the P7 and the Pusen 9.2F. All scopes had illuminance skew, except the V3. All scopes had a warm color temperature. CONCLUSION: Illumination properties vary between ureteroscopes in an enclosed cavity in saline, and differs at center vs 45° and 90° positions within scopes. Peripheral illuminance drop can be as high as - 92%, which is undesirable. This may affect the choice of ureteroscope and light brightness settings used in surgery by urologists.


Assuntos
Desenho de Equipamento , Rim , Iluminação , Ureteroscópios , Modelos Anatômicos , Humanos
2.
World J Urol ; 42(1): 292, 2024 May 04.
Artigo em Inglês | MEDLINE | ID: mdl-38704492

RESUMO

PURPOSE: During endoscopic stone surgery, Holmium:YAG (Ho:YAG) and Thulium Fiber Laser (TFL) technologies allow to pulverize urinary stones into fine particles, ie DUST. Yet, currently there is no consensus on the exact definition of DUST. This review aimed to define stone DUST and Clinically Insignificant Residual Fragments (CIRF). METHODS: Embase, MEDLINE (PubMed) and Cochrane databases were searched for both in vitro and in vivo articles relating to DUST and CIRF definitions, in November 2023, using keyword combinations: "dust", "stones", "urinary calculi", "urolithiasis", "residual fragments", "dusting", "fragments", "lasers" and "clinical insignificant residual fragments". RESULTS: DUST relates to the fine pulverization of urinary stones, defined in vitro as particles spontaneously floating with a sedimentation duration ≥ 2 sec and suited for aspiration through a 3.6Fr-working channel (WC) of a flexible ureteroscope (FURS). Generally, an upper size limit of 250 µm seems to agree with the definition of DUST. Ho:YAG with and without "Moses Technology", TFL and the recent pulsed-Thulium:YAG (pTm:YAG) can produce DUST, but no perioperative technology can currently measure DUST size. The TFL and pTm:YAG achieve better dusting compared to Ho:YAG. CIRF relates to residual fragments (RF) that are not associated with imminent stone-related events: loin pain, acute renal colic, medical or interventional retreatment. CIRF size definition has decreased from older studies based on Shock Wave Lithotripsy (SWL) (≤ 4 mm) to more recent studies based on FURS (≤ 2 mm) and Percutaneous Nephrolithotomy(PCNL) (≤ 4 mm). RF ≤ 2 mm are associated with lower stone recurrence, regrowth and clinical events rates. While CIRF should be evaluated postoperatively using Non-Contrast Computed Tomography(NCCT), there is no consensus on the best diagnostic modality to assess the presence and quantity of DUST. CONCLUSION: DUST and CIRF refer to independent entities. DUST is defined in vitro by a stone particle size criteria of 250 µm, translating clinically as particles able to be fully aspirated through a 3.6Fr-WC without blockage. CIRF relates to ≤ 2 RF on postoperative NCCT.


Assuntos
Ureteroscopia , Cálculos Urinários , Humanos , Cálculos Urinários/terapia , Litotripsia a Laser/métodos , Poeira , Tamanho da Partícula
3.
World J Urol ; 42(1): 294, 2024 May 05.
Artigo em Inglês | MEDLINE | ID: mdl-38704777

RESUMO

PURPOSE: To date, no study has evaluated effects of varying brightness settings on image quality from flexible ureteroscopes submerged in saline. The aim was to evaluate blackout and whiteout occurrences in an in-vitro kidney calyx model. MATERIAL AND METHODS: We evaluated a series of contemporary flexible ureteroscopes including the Storz Flex-Xc and Flex-X2s, Olympus V3 and P7, Pusen 7.5F and 9.2F, as well as OTU WiScope using a 3D-printed enclosed pink in-vitro kidney calyx model submerged in saline. Endoscopic images were captured with ureteroscope tip placed at 5 mm,10 mm and 20 mm distances. The complete range of brightness settings and video capture modes were evaluated for each scope. Distribution of brightness on a grayscale histogram of images was analyzed (scale range 0 to 255). Blackout and whiteout were defined as median histogram ranges from 0 to 35 and 220 to 255, respectively (monitor image too dark or too bright for the human eye, respectively). RESULTS: Blackout occurred with the P7, Pusen 7.5F, 9.2F and WiScope at all distances, and V3 at 20 mm - with lowest brightness settings. Whiteout occurred with Flex-X2s, V3 and P7 at 5 mm and 10 mm, as well as with V3 and P7 at 20 mm - mostly with highest brightness settings. The Flex-Xc had neither blackout nor whiteout at all settings and distances. CONCLUSION: Blackout or whiteout of images is an undesirable property that was found for several scopes, possibly impacting diagnostic and therapeutic purposes during ureteroscopy. These observations form a guide to impact a urologist's choice of instruments and settings.


Assuntos
Ureteroscópios , Ureteroscopia , Humanos , Desenho de Equipamento , Iluminação , Maleabilidade , Cálices Renais
4.
World J Urol ; 42(1): 233, 2024 Apr 13.
Artigo em Inglês | MEDLINE | ID: mdl-38613608

RESUMO

PURPOSE: This study aimed to evaluate the ability of Kidney Stone Calculator (KSC), a flexible ureteroscopy surgical planning software, to predict the lithotripsy duration with both holmium:YAG (Ho:YAG) and thulium fiber laser (TFL). METHODS: A multicenter prospective study was conducted from January 2020 to April 2023. Patients with kidney or ureteral stones confirmed at non-contrast computed tomography and treated by flexible ureteroscopy with laser lithotripsy were enrolled. "Kidney Stone Calculator" provided stone volume and subsequent lithotripsy duration estimation using three-dimensional segmentation of the stone on computed tomography and the graphical user interface for laser settings. The primary endpoint was the quantitative and qualitative comparison between estimated and effective lithotripsy durations. Secondary endpoints included subgroup analysis (Ho:YAG-TFL) of differences between estimated and effective lithotripsy durations and intraoperative outcomes. Multivariate analysis assessed the association between pre- and intraoperative variables and these differences according to laser source. RESULTS: 89 patients were included in this study, 43 and 46 in Ho:YAG and TFL groups, respectively. No significant difference was found between estimated and effective lithotripsy durations (27.37 vs 28.36 min, p = 0.43) with a significant correlation (r = + 0.89, p < 0.001). Among groups, this difference did not differ (p = 0.68 and 0.07, respectively), with a higher correlation between estimated and effective lithotripsy durations for TFL compared to Ho:YAG (r = + 0.95, p < 0.001 vs r = + 0.81, p < 0.001, respectively). At multivariate analysis, the difference was correlated with preoperative (volume > 2000 mm3 (Ho:YAG), 500-750 mm3 SV and calyceal diverticulum (TFL)), operative (fragmentation setting (p > 0.001), and basket utilization (p = 0.05) (Ho:YAG)) variables. CONCLUSION: KSC is a reliable tool for predicting the lithotripsy duration estimation during flexible ureteroscopy for both Ho:YAG and TFL. However, some variables not including laser source may lead to underestimating this estimation.


Assuntos
Cálculos Renais , Litotripsia , Cálculos Ureterais , Humanos , Hólmio , Túlio , Ureteroscopia , Estudos Prospectivos , Cálculos Renais/cirurgia , Lasers
5.
Eur Urol Open Sci ; 63: 89-95, 2024 May.
Artigo em Inglês | MEDLINE | ID: mdl-38585592

RESUMO

Background and objective: Data regarding open conversion (OC) during minimally invasive surgery (MIS) for renal tumors are reported from big databases, without precise description of the reason and management of OC. The objective of this study was to describe the rate, reasons, and perioperative outcomes of OC in a cohort of patients who underwent MIS for renal tumor initially. The secondary objective was to find the factors associated with OC. Methods: Between 2008 and 2022, of the 8566 patients included in the UroCCR project prospective database (NCT03293563), who underwent laparoscopic or robot-assisted minimally invasive partial (MIPN) or radical (MIRN) nephrectomy, 163 experienced OC. Each center was contacted to enlighten the context of OC: "emergency OC" implied an immediate life-threatening situation not reasonably manageable with MIS, otherwise "elective OC". To evaluate the predictive factors of OC, a 2:1 paired cohort on the UroCCR database was used. Key findings and limitations: The incidence rate of OC was 1.9% for all cases of MIS, 2.9% for MIRN, and 1.4% for MIPN. OC procedures were mostly elective (82.2%). The main reason for OC was a failure to progress due to anatomical difficulties (42.9%). Five patients (3.1%) died within 90 d after surgery. Increased body mass index (BMI; odds ratio [OR]: 1.05, 95% confidence interval [CI]: 1.01-1.09, p = 0.009) and cT stage (OR: 2.22, 95% CI: 1.24-4.25, p = 0.008) were independent predictive factors of OC. Conclusions and clinical implications: In MIS for renal tumors, OC was a rare event (1.9%), caused by various situations, leading to impaired perioperative outcomes. Emergency OC occurred once every 300 procedures. Increased BMI and cT stage were independent predictive factors of OC. Patient summary: The incidence rate of open conversion (OC) in minimally invasive surgery for renal tumors is low. Only 20% of OC procedures occur in case of emergency, and others are caused by various situations. Increased body mass index and cT stage were independent predictive factors of OC.

6.
World J Urol ; 42(1): 188, 2024 Mar 23.
Artigo em Inglês | MEDLINE | ID: mdl-38520528

RESUMO

PURPOSE: Artifacts from poor ureteroscopes' light design with shadowing and dark areas in the field of view have been reported. The aim was to quantify effects of light obstruction in a kidney calyx model. METHODS: We evaluated a series of contemporary flexible ureteroscopes including the Storz Flex-Xc and Flex-X2s, Olympus V3 and P7, Pusen 7.5F and 9.2F, as well as OTU Wiscope using an enclosed 3D-printed pink in vitro kidney calyx model submerged in saline, where the field of light was intentionally partially obstructed alternatively at 12, 3, 6, and 9 o'clock. A color spectrometer was used for illuminance measurements at a 45° opening position in the background of the model. RESULTS: Overall and mean background illuminance for each obstructive situation were significantly different between scopes for both 50% and 100% brightness settings (ANOVA p < 0.001). At 50% brightness setting, almost all scopes had their highest and lowest background illuminance with the 6 o'clock and 3 o'clock obstructive situation, respectively. At 100% brightness setting, these became 6 o'clock and 12 o'clock obstructive situations. Considering each obstructive situation individually, the Flex-Xc was consistently the scope with highest background illuminance and the Pusen 7.5F the lowest. Background illuminance for each obstructive situation varied significantly for each scope individually, with the greatest range of variability for Pusen 7.5F and V3. CONCLUSIONS: Illuminance performance of ureteroscopes within an obstructed calyx model differ significantly for various obstructive situations. Urologists should be aware of this to help guide their choice of ureteroscope.


Assuntos
Iluminação , Ureteroscópios , Humanos , Desenho de Equipamento , Urologistas , Equipamentos Descartáveis , Ureteroscopia
7.
Fr J Urol ; 34(4): 102609, 2024 Mar 08.
Artigo em Inglês | MEDLINE | ID: mdl-38460936

RESUMO

Angiosarcoma is a rare malignancy derived from endothelial cells, which behaves aggressively. Primary angiosarcoma of the kidney is even rarer, and its clinical and radiological presentations do not differ from clear cells renal cell carcinoma (ccRCC). Management protocols are not standardized, although nephrectomy is usually performed. Subsequent treatments (chemotherapy, radiotherapy, and lately, targeted therapies) vary considerably. Herein, we report the case of a middle-aged patient harboring primary angiosarcoma of the left kidney and discuss its presentation and management in light of current guidelines. The case is described for its rarity and masquerading nature.

8.
J Endourol ; 38(5): 488-498, 2024 May.
Artigo em Inglês | MEDLINE | ID: mdl-38429915

RESUMO

Urothelial carcinoma (UC) affecting the urinary tract is recognized as one of the prevalent types of cancer, ranking fifth in terms of incidence. However, it is important to note that upper tract urothelial carcinoma (UTUC), in comparison to bladder cancer, is relatively uncommon and represents a minority of UC cases, accounting for ∼5% to 10%. It is worth emphasizing that as much as 60% of UTUC cases are invasive at the time of diagnosis, in contrast to 25% of bladder cancer cases. These statistics highlight the urgent need for well-designed, multidisciplinary strategies to guide optimal management for this vulnerable patient population, aiming to control treatment variability and minimize the risks associated with undertreatment and overtreatment. Methods: In this study, we present a comparative analysis of the contemporary guidelines on UTUC management from the European Association of Urology (EAU) and the American Urological Association/Society of Urologic Oncology (AUA/SUO) released in 2023. Our objective is to contrast the different guidelines and examine the evidence on which their recommendations are based. Results: By thoroughly reviewing the guidelines from both organizations, we observed substantial agreement on the management of UTUC. However, we also identified noticeable differences among these guidelines, leading to a wide range of recommendations. These disparities might stem from variations in clinical practices, regional preferences, and the availability of resources. It is crucial to acknowledge that both the EAU and AUA/SUO base their guidelines on the latest scientific evidence and expert consensus within their respective regions. Conclusions: These findings underscore the importance of ongoing collaboration, knowledge exchange, and harmonization of guidelines to improve the standard of care for UTUC globally. Future research should focus on identifying areas of consensus and bridging the gaps between different international guidelines to enhance the management outcomes for this challenging disease.


Assuntos
Guias de Prática Clínica como Assunto , Humanos , Europa (Continente) , Estados Unidos , Carcinoma de Células de Transição/terapia , Carcinoma de Células de Transição/diagnóstico , Carcinoma de Células de Transição/patologia , Neoplasias Urológicas/terapia
11.
World J Urol ; 42(1): 57, 2024 Jan 27.
Artigo em Inglês | MEDLINE | ID: mdl-38280001

RESUMO

INTRODUCTION: Both Holmium:yttrium-aluminium-garnet (Ho:YAG) laser and Thulium Fiber Laser (TFL) can effectively treat all urinary stone types. This in vitro study evaluated the ablation volume per pulse (AVP) and required energy needed to ablate 1mm3 (RE, J/mm3) of various stone types at different laser settings with TFL. METHODS: 272-µm core-diameter laser fibers (Boston Scientific©) were connected to a 50 Watts TFL generator (IPG®). An experimental setup immerged human stones of calcium oxalate monohydrate (COM), uric acid (UA), and cystine (CYS) with a single pulse lasing emission (0.5/0.8/1 J), in contact mode. Stones were dried out before three-dimensional scanning to measure AVP and deduce from the pulse energy (PE) and AVP the RE. A direct comparison with known Ho:YAG's AVP and RE was then carried out. RESULTS: AVP for COM stones was significantly greater than those for CYS stones and similar to UA stones (p = 0.02 and p = 0.06, respectively). If AVP increased with PE against COM and UA stones, AVP decreased against CYS stones. 1 J PE resulted in a threefold lower RE compared with other PE for COM stones. On the contrary, RE for CYS increased with PE, whereas PE did not had influence on RE for UA. TFL was associated with greater AVP for COM, but lower for UA and CYS stones compared to Ho:YAG laser. CONCLUSION: This in vitro study firstly describes the ablation volume per pulse and required energy to treat a cubic millimeter of three frequent human stone types, and suggest TFL could not be suited for cystine. Therefore, stone composition could be considered when choosing the laser source for lithotripsy.


Assuntos
Terapia a Laser , Lasers de Estado Sólido , Litotripsia a Laser , Nefrolitíase , Cálculos Urinários , Humanos , Litotripsia a Laser/métodos , Túlio , Cistina , Cálculos Urinários/cirurgia , Lasers de Estado Sólido/uso terapêutico , Hólmio
12.
World J Urol ; 41(12): 3765-3771, 2023 Dec.
Artigo em Inglês | MEDLINE | ID: mdl-37833547

RESUMO

BACKGROUND AND OBJECTIVE: We aimed to evaluate the concordance between the pre-settings ranges of thulium fibre laser (TFL) (Coloplast TFL Drive, Denmark) with easy-to-use graphical user interface and the laser settings used by a high-volume endo-urologist during surgical procedures. MATERIALS AND METHODS: In October 2022, we prospectively collected data of 67 patients who underwent TFL Drive (Coloplast, Denmark) for the management of urinary stones. Urothelial tumour (upper tract urinary cancer (UTUC) and bladder) 200 and 150 µm laser fibres were used for procedures. Stones characteristics (size and density) tumours and stenosis localizations, laser-on time (LOT), and laser settings were recorded. We also assessed the ablation speed (mm3/s), laser power (W), and Joules/mm3 values for each lithotripsy. RESULTS: A total 67 patients took part in the study. Median age was 52 (15-81) years. 55 (82%), 8 (12%), and 4 (6%) patients presented urinary stones, urothelial tumour, and stenosis, respectively. Median stone volume was 438 (36-6027) mm3 and median density was 988 (376-2000) HU. Median pulse energy was 0.6 (0.3-1.2), 0.8 (0.5-1) and 1 J for urinary stones, urothelial tumour and stenosis respectably. Endoscopically stone-free rate was 89%. Graphical user interface and surgeon accordance with the safety range were observed in 93.2%, 100% and 100% for urinary stones, UTUC and stenosis, respectively. CONCLUSION: During endoscopic procedures for urinary stones treatment, it is frequently needed to change laser parameters. These new TFL and GUI technology parameters remained in the pre-set security range in 94.1% of procedures.


Assuntos
Lasers de Estado Sólido , Litotripsia a Laser , Neoplasias , Cálculos Urinários , Humanos , Pessoa de Meia-Idade , Túlio , Litotripsia a Laser/métodos , Constrição Patológica , Cálculos Urinários/cirurgia , Lasers de Estado Sólido/uso terapêutico
13.
World J Urol ; 41(11): 3345-3353, 2023 Nov.
Artigo em Inglês | MEDLINE | ID: mdl-37728745

RESUMO

PURPOSE: Flexible ureteroscopy (fURS) is steadily gaining popularity in the management of renal calculi, including those located in the lower pole (LP). Due to difficulty in accessing to the LP of kidney in minority of cases with fURS and reports of lower stone-free rate (SFR), it is still considered as a challenge in selected cases. The purpose of the review was to analyze the various aspects of fURS for LP stones. METHODS: An extensive review of the recent literature was done including different factors such as anatomy, preoperative stenting, stone size, flexible scopes, types of lasers, laser fibers, suction, relocation, stone-free rates, and complications. RESULTS: The significance of various lower pole anatomical measurements remain a subject of debate and requires standardization. Recent improvements in fURS such as single-use digital scopes with better vision and flexibility, high power laser, thulium fiber laser, smaller laser fiber, and accessories have significantly contributed to make flexible ureteroscopy  more effective and safer in the management of LP stone. The utilization of thulium fiber lasers in conjunction with various suction devices is being recognized and can significantly improve SFR. CONCLUSIONS: With the significant advancement of various aspects of fURS, this treatment modality has shown remarkable efficacy and gaining widespread acceptance in management of LP kidney stones. These developments have made the fURS of LP stones less challenging.


Assuntos
Cálculos Renais , Ureteroscopia , Humanos , Túlio , Ureteroscópios , Cálculos Renais/cirurgia , Rim , Resultado do Tratamento
14.
J Clin Med ; 12(15)2023 Jul 26.
Artigo em Inglês | MEDLINE | ID: mdl-37568309

RESUMO

INTRODUCTION: Thulium lasers (TLs), namely the Thulium fiber laser (TFL) and the Thulium:YAG (Tm:YAG), are being increasingly adopted for the conservative treatment of upper urinary tract urothelial carcinoma (UTUC). However, to date, the real clinical impact of TLs on UTUC management remains not well-characterized. We performed a review of the literature to summarize the current evidence on TLs for UTUC treatment. MATERIALS AND METHODS: We performed a systematic review in January 2023 using the Embase and Medline online databases, according to the PRISMA recommendations and using the PICO criteria. Outcomes of interest were: (i) to assess the safety and feasibility of TLs in the treatment of UTUC, and (ii) to evaluate the oncological outcomes in terms of tumor recurrence and conservative treatment failure. Moreover, we described TL characteristics and its interaction with soft tissue. RESULTS: a total of 458 articles were screened, and six full texts including 273 patients were identified. All the included studies were retrospective series. Mean patient age ranged from 66 to 73 years. The indication of a conservative treatment was elective and imperative in 21.7-85% and 15-76% of cases, respectively. Laser power settings varied from 5 to 50 W. No intraoperative complications were reported, and all the procedures were successfully performed. The tumor recurrence rate was 17.7-44%, and the indication to radical nephroureterectomy was 3.7-44% during a follow-up of 6-50 months. Most of the postoperative complications were mild and transient, and ureteral strictures were reported in two studies. Major limitations were the retrospective nature of the studies, the small sample sizes, and the short follow-up. CONCLUSIONS: TL is an effective and safe technology for endoscopic UTUC treatment. However, current available literature lacks prospective and multicentric studies with large population sizes and long-term follow-up.

15.
World J Urol ; 41(8): 2119-2125, 2023 Aug.
Artigo em Inglês | MEDLINE | ID: mdl-37414942

RESUMO

INTRODUCTION: Holmium:yttrium-aluminium-garnet (Ho:YAG) and thulium fiber (TFL) lasers are currently the two laser sources recommended for endocorporeal laser lithotripsy (ELL). Recently, the pulsed-thulium:YAG (Tm:YAG) laser was also proposed for ELL, as an answer to both Ho:YAG and TFL limitations. We aimed to evaluate the efficiency, safety, and laser settings of Tm:YAG laser in ELL during retrograde intrarenal surgery (RIRS). METHODS: A prospective study of the first 25 patients with ureteral and renal stones who underwent RIRS using the Thulio (pulsed-Tm:YAG, Dornier©, Germany) was performed in a single center. 272 µm laser fibers were used. Stone size, stone density, laser-on time (LOT) and laser settings were recorded. We also assessed the ablation speed (mm3/s), Joules/mm3 and laser power (W) values for each procedure. Postoperative results, such as stone-free rate (SFR) and zero fragments rate (ZFR) were also recorded. RESULTS: A total of 25 patients were analyzed (Table 1). The median (IQR) age was 55 (44-72) years old. Median (IQR) stone volume was 2849 (916-9153)mm3. Median (IQR) stone density was 1000 (600-1174)HU. Median (IQR) pulse energy, pulse rate and total power were 0.6 (0.6-0,8)J, 15(15-20)Hz and 12(9-16)W, respectively. All procedures used "Captive Fragmenting" pulse modulation (Table 2). The median (IQR) J/mm3 was 14,8 (6-21). The median (IQR) ablation rate was 0,75 (0,46-2)mm3/s. One postoperative complications occurred (streinstrasse). SFR and ZFR were 95% and 55%, respectively. CONCLUSION: The pulsed-Tm:YAG laser is a safe and effective laser source for lithotripsy during RIRS, using low pulse energy and low pulse frequency.


Assuntos
Cálculos Renais , Lasers de Estado Sólido , Litotripsia a Laser , Humanos , Pessoa de Meia-Idade , Idoso , Lasers de Estado Sólido/uso terapêutico , Túlio/uso terapêutico , Estudos Prospectivos , Litotripsia a Laser/métodos , Cálculos Renais/cirurgia , Hólmio
16.
Curr Opin Urol ; 33(4): 308-317, 2023 07 01.
Artigo em Inglês | MEDLINE | ID: mdl-37140545

RESUMO

PURPOSE OF REVIEW: Retrograde intra renal surgery (RIRS) with laser lithotripsy represents the gold-standard to treat renal stones up to 20 mm. Controlling intraoperative parameters such as intrarenal pressure (IRP) and temperature (IRT) is mandatory to avoid complications. This article reviews advances in IRP and IRT over the last 2 years. RECENT FINDINGS: We conducted a PubMed/Embase search and reviewed publications that include temperature and pressure during RIRS. Thirty-four articles have been published which met the inclusion criteria. Regarding IRP, a consensus has emerged to control IRP during RIRS, in order to avoid (barotraumatic and septic) complications. Several monitoring devices are under evaluation but none of them are clinically approved for RIRS. Ureteral access sheath, low irrigation pressure and occupied working channel help to maintain a low IRP. Robotic systems and suction devices would improve IRP intraoperative management and monitoring. IRT determinants are the irrigation flow and laser settings. Low power settings(<20 W) with minimal irrigation flow (5-10 ml/min) are sufficient to maintain low IRT and allows continuous laser activation. SUMMARY: Recent evidence suggests that IRP and IRT are closely related. IRP depends on inflow and outflow rates. Continuous monitoring would help to avoid surgical and infectious complications. IRT depends on the laser settings and the irrigation flow.


Assuntos
Cálculos Renais , Litotripsia a Laser , Ureter , Humanos , Temperatura , Rim/cirurgia , Cálculos Renais/cirurgia
17.
J Clin Med ; 12(8)2023 Apr 11.
Artigo em Inglês | MEDLINE | ID: mdl-37109152

RESUMO

OBJECTIVE: The aim of this review is to summarize the existing suction systems in flexible ureteroscopy (fURS) and to evaluate their effectiveness and safety. METHODS: A narrative review was performed using the Pubmed and Web of Science Core Collection (WoSCC) databases. Additionally, we conducted a search on the Twitter platform. Studies including suctions systems in fURS were included. Editorials, letters and studies reporting intervention with semirigid ureteroscopy, PCNL and mPCNL were excluded. RESULTS: A total of 12 studies were included in this review. These studies comprised one in vitro study, one ex vivo study, one experimental study and eight cohort studies. The Pubmed and WoSCC searches identified three suction techniques (Irrigation/Suctioning system with control of pressure, suction ureteral access sheath (sUAS) and direct in scope suction (DISS)), and the Twitter search identified four of them. The overall results showed that suction is an effective and safe technique that improves stone-free rates, reduces operative time and limits complication rates after fURS. CONCLUSIONS: The use of suctioning during common endourological procedures has been shown to improve safety and efficacy in several indications. However, randomized controlled trials are needed to confirm this.

19.
J Endourol ; 36(2): 251-258, 2022 02.
Artigo em Inglês | MEDLINE | ID: mdl-34409842

RESUMO

Objective: To assess the fiber-tissue interaction through ablation, coagulation, and carbonization characteristics of the Ho:YAG laser and super pulsed thulium fiber laser (TFL) in a nonperfused porcine kidney model. To assess the degradation of laser fibers during soft tissue treatment. Methods: A 50 W TFL generator was compared with a 120 W Ho:YAG laser. The laser settings that can be set identically between the two lasers (pulse energy and frequency), and clinically relevant for prostate laser enucleation, were identified and used for tissue incisions on fresh nonfrozen porcine kidneys. For each parameter, the short, medium, and long pulse durations for the Ho:YAG generator and the different peak powers 150, 250, and 500 W for the TFL generator were also tested. Laser incisions were performed with 550 µm stripped laser fiber fixed on a robotic arm at a distance of 0.1 mm with the tissue surface and at a constant speed of 10 mm/s. Histologic analysis was then performed, evaluating incision shape, incision depth and width, axial coagulation depth, and presence of carbonization. Degradation of the laser fiber was defined as reduction of laser fiber tip length after laser activation. Results: Incision depths and areas of coagulation were greater with the Ho:YAG laser than those with the TFL. Although no carbonization zone was found with the Ho:YAG laser, this was constant with the TFL. Although a fiber tip degradation was constantly observed with Ho:YAG laser, except in the case of a long pulse duration and low pulse energy (0.2 J), this was not the case with TFL. Conclusion: TFL appears to be an efficient alternative to Ho:YAG laser for soft tissue surgery. The histologic analysis found greater tissue penetration with the Ho:YAG laser and different coagulation properties between the two lasers. These results need to be investigated in vivo to assess the clinical impact of these differences and find the optimal settings for laser prostate enucleation.


Assuntos
Terapia a Laser , Lasers de Estado Sólido , Litotripsia a Laser , Animais , Hólmio , Terapia a Laser/métodos , Lasers de Estado Sólido/uso terapêutico , Litotripsia a Laser/métodos , Masculino , Próstata , Suínos , Túlio
20.
World J Urol ; 39(9): 3607-3614, 2021 Sep.
Artigo em Inglês | MEDLINE | ID: mdl-33779821

RESUMO

INTRODUCTION: Kidney Stone Calculator (KSC) is a free, three-dimensional (3D) planning software for flexible ureteroscopy(fURS) with Holmium:YAG(Ho:YAG) endocorporeal lithotripsy (EL). KSC provides the stone volume (SV) and expected duration of lithotripsy (ExDL) estimations based on non-enhanced-CT scan (NECT) DICOM series. We aimed to provide a preclinical and clinical evaluation of KSC. PATIENTS AND METHODS: A preclinical evaluation measured the SV by three operators (resident, endourology expert and research engineer) among 17 NECT cases. Between January and March 2020, a multicentric, prospective, observational double-blind clinical evaluation was conducted in patients presenting with renal stones treated with Ho:YAG-EL during fURS and preoperative NECT. Demographic and surgical data were collected. The primary endpoint was a significant median difference between ExDL and EffectiveDL (EfDL). Second, efficiency (J/mm3) and efficacy (mm3/min) ratios were calculated. RESULTS: The preclinical evaluation showed no significant difference in the SV measurements among operators (p > 0.05). Pearson and Kendall coefficients of 0.99 and 0.98, respectively, were found. Twenty-six patients were included in the clinical evaluation, with a median age of 55 years. In 66% of cases, there was a single stone located in the lower pole, with a density > 1000 Hounsfield Unit observed in 42% and 85% of cases. A 14% [Q1-Q3 (5.4-24.8); p = 0.36] median difference between ExDL and EfDL was noted, which was greater in the case of lower pole stones with no possible relocation (p = 0.008). Median values of 17.6 J/mm3 and 0.4 (0.32-0.56) mm3/s EL were also noted. CONCLUSIONS: Kidney Stone Calculator is a reproducible and accurate software that allows for an estimation of the stone burden and provides an ExDL for URSf. Defining the influencing factors of EL will improve its ExDL.


Assuntos
Cálculos Renais/cirurgia , Lasers de Estado Sólido/uso terapêutico , Planejamento de Assistência ao Paciente , Software , Ureteroscopia/métodos , Adulto , Idoso , Método Duplo-Cego , Feminino , Humanos , Masculino , Pessoa de Meia-Idade , Projetos Piloto , Estudos Prospectivos
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA